Τετάρτη 31 Μαΐου 2017

Amperometric detector for gas chromatography based on a silica sol-gel solid electrolyte

Publication date: 1 November 2017
Source:Talanta, Volume 174
Author(s): William H. Steinecker, Krzysztof Miecznikowski, Pawel J. Kulesza, Zechariah D. Sandlin, James A. Cox
An electrochemical cell comprising a silica sol-gel solid electrolyte, a working electrode that protrudes into a gas phase, and reference and counter electrodes that contact the solid electrolyte comprises an amperometric detector for gas chromatography. Under potentiostatic conditions, a current related to the concentration of an analyte in the gas phase is produced by its oxidation at the three-phase boundary among the sol-gel, working electrode, and the gas phase. The sol-gel is processed to contain an electrolyte that also serves as a humidistat to maintain a constant water activity even in the presence the gas chromatographic mobile phase. Response was demonstrated toward a diverse set of analytes, namely hydrogen, 1,2-ethandithiol, phenol, p-cresol, and thioanisole. Using flow injection amperometry of hydrogen with He as the carrier gas, 90% of the steady-state current was achieved in < 1s at a flow rate of 20mLmin−1. A separation of 1,2-ethandithiol, phenol, p-cresol, and thioanisole at a 2.2mLmin−1 flow rate was achieved with respective detection limits (k = 3 criterion) of 4, 1, 3, and 70 ppmv when the working electrode potential was 800mV.

Graphical abstract

image


http://ift.tt/2rUeCHl