Τετάρτη 6 Οκτωβρίου 2021

Parathyroid Gland Autofluorescence Characteristics in Patients With Primary Hyperparathyroidism

xlomafota13 shared this article with you from Inoreader

Objective

Near-infrared imaging for intraoperative parathyroid gland (PG) detection has recently commanded significant attention. The PTeye (Medtronic, Minneapolis, MN) is a probe-based system for near-infrared autofluorescent evaluation of PGs. This study was designed to evaluate the capabilities of the PTeye in the setting of surgery for primary hyperparathyroidism.

Study Design

Prospective, Cohort study.

Methods

This single-institution, prospective cohort study included all patients undergoing parathyroidectomy for primary hyperparathyroidism with presumed single gland disease from June 2020 to December 2020. Absolute intensity and intensity ratios, with the thyroid as the control tissue, were obtained for the adenoma, ipsilateral normal PG, and adjacent tissue. The ability of the PTeye to function when not in direct contact with tissue was measured.

Results

Twenty-two patients were included. The median intensity ratio for the in situ adenomas was 4.38 (interquartile range [IQR]: 2.03–5.87), ipsilateral normal PGs 6.17 (IQR: 3.83–7.67), strap muscle 0.47 (IQR: 0.30–0.60), and fat 0.20 (IQR: 0.17–0.47). All normal PGs and 21/22 adenomas demonstrated autofluorescence above the detection threshold. The PTeye functioned at a maximum distance of separation of 10 mm through saline medium and 6 mm through clear solid medium.

Conclusion

This study confirms the PTeye's ability to recognize PGs with a high degree of precision. The device was found to function properly even with the probe not in direct contact with the tissue. Although adenomatous PGs appear to demonstrate altered autofluorescent properties from normal PGs, additional research is required to determine if these differences are clinically useful.

Level of Evidence

3 Laryngoscope, 2021

View on the web

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου

Δημοφιλείς αναρτήσεις